|
|
![]() |
|
|
|
| Kumar RK, Nair AC. Coil Occlusion of the Large Patent Ductus Arteriosus. Images Paediatr Cardiol 2008;34:8-26 | |
| Amrita Institute of Medical Sciences and Research Centre, Kochi, Kerala, India |
| Heart Catheterization | Patent ductus arteriosus | |
Initial reports of the use of Gianturco coils for the patent ductus arteriosus were published in early 1990s.1 Coil occlusion is now almost universally established as a simple, safe and effective technique for occlusion of the small patent ductus arteriosus (PDA) that measures less than 3 mm in diameter at its narrowest point (usually at the site of insertion of the duct in the pulmonary artery).2,3 Coil occlusion of larger ducts is technically challenging because of a greater tendency for coil embolization.4 Various modifications in technique have been suggested to reduce the risk of embolization. They include the use of detachable coils,5 deployment of thicker (0.052 inch) coils,6 simultaneous deployment of two or more coils,7 snare-assisted delivery,8 and bioptome assisted-delivery.9-11 Occlusive devices such as the Amplatzer Duct Occluder (ADO) overcome many limitations of coils for closure of large PDAs and allow for better control and safety. Most institutions now prefer occlusive devices for PDAs that are >3 mm. Devices are however considerably more expensive than coils, and in many developing countries device closure costs substantially more than surgical closure. The bioptome-assisted coil occlusion technique has emerged as a less-expensive alternative to the ADO.10-13 With careful attention to case selection and technique it is possible to coil occlude the majority of ducts. Further, in specific instances, such as in selected small infants, coil occlusion may have an advantage over the Amplatzer duct occluder. This review will describe case selection strategies and the coil occlusion technique for ducts that are relatively large (? 3mm at the narrowest point).
Anatomic and physiologic considerations
The PDA is typically an asymmetric, truncated cone with considerable
variations in its size, shape and attachment to the aorta (Fig. 1). The
narrowest part of the duct is typically close to the pulmonary arterial
end of the duct. This is perhaps because of the tendency of the duct to
close from its pulmonary artery (PA) end early after birth. Variations
in the size of the ampulla often determine suitability for coil occlusion.14

The ampulla in the majority of ducts is generous enough to accommodate
coils of appropriate size for occluding the PA end. In some instances,
however, the ampulla is shallow. Rarely, it is absent altogether. The broader
aortic end of the ampulla typically originates from the leftward aspect
of the aorta. It is therefore often necessary to undertake the aortogram
in the right anterior view as well as the conventional lateral view.
Unless the pulmonary vascular resistance is considerably elevated, the PDA shunts continuously from aorta to PA. The duct size and relative resistances of the systemic and pulmonary artery circuits determine the extent of flow reversal that occurs in the proximal descending aorta. From the standpoint of coil occlusion, flow reversal is a useful phenomenon because it can be used to for angiographic definition (see below) and for deposition of coils. The hemodynamic significance of the duct is determined by size, length (longer ducts are likely to offer greater resistance to flows) and age at presentation.
Clinical cvaluation, chest X-ray and ECG
A good clinical evaluation provides valuable clues on the likely size
of the duct and pulmonary blood flow. Increased precordial activity, bounding
pulses and a murmur that is grade III or louder all suggest that the duct
is likely to be large with a significant left to right shunt. The chest
X-ray in these patients typically reveals increased pulmonary blood flow
and cardiac enlargement with a prominent aorta and ECG reveals prominent
left ventricular forces with prominent q waves in lateral chest leads.
Clinical, X-ray and ECG features of elevated pulmonary vascular resistance
may preclude the possibility of coil occlusion.
Echocardiography
A detailed echocardiographic evaluation of the duct anatomy
and physiology is a must for all patients undergoing coil occlusion. Excellent
definition of anatomy is feasible in almost all infants and children and
in many adults.15,16 The duct diameter should be measured in
a high parasternal long-axial view at the point where it opens into the
pulmonary artery (Fig. 2, 3). The measurement
must be made using the zoomed two-dimensional echocardiographic image (Fig.
2) and not the width of the color Doppler jet. Subtle adjustments in the
transducer position and angulations are required for precise definition
of the PDA at its pulmonary artery insertion.


The ductal ampulla is considered adequate if its maximal dimension along
the long axis is greater than twice the measured ductal diameter. (Fig.
4) Essentially, one needs to visualize whether a coil that is large enough
to occlude the duct can occupy the ampulla without protrusion into the
aorta. It is a good idea to draw a picture of the echo anatomy (Fig. 4)
in the report and plan the strategy for coil occlusion based on the echocardiographic
anatomy.

Echocardiography provides many clues about the physiologic significance
of the duct. A large duct with increased pulmonary blood flow is suggested
by a left atrial and left ventricular enlargement and flow reversal in
the descending thoracic and abdominal aorta. The Doppler gradients across
such a duct at end diastole are typically low (< 30 mm Hg). In our experience,
peak systolic Doppler gradients do not correlate well with duct size. Major
elevation of pulmonary vascular resistance is suggested by low velocities
in both directions across the PDA together with absence of flow reversal
in the descending aorta. These ducts are typically very large and coil
occlusion is often not an acceptable option.
In addition to evaluation of the PDA, the origins of the main branch pulmonary arteries should be carefully inspected for stenosis at their origins. Any internal inconsistencies between duct size estimation and hemodynamic correlates (above) should prompt reassessment of size through repeat measurements. For example if the duct is measured as 2 mm in a child who has a large shunt that is clinically obvious with echocardiographic features of a large shunt, there is a distinct possibility of an error in the measurement.
Case selection
The decision on the closure strategy for PDA is determined by the following
considerations:
Hemodynamic evaluations
Standard hemodynamic measures should be obtained. In the event that
the pulmonary vascular resistance is elevated and there are doubts on whether
the PDA closure would reverse the pulmonary artery hypertension, it may
be wise to balloon occlude the duct and measure the pulmonary artery pressures.
Balloon occlusion can be accomplished by a balloon end-hole catheter (7F
Swan Ganz catheter is well suited for this purpose) passed from the venous
route into the aorta. The inflated balloon is then pulled back into the
duct. This may necessitate an additional venous access. Alternatively the
balloon catheter may be passed via a larger long sheath and the PA pressure
can be measured from the side arm of the sheath. The systemic arterial
pressure should be measured simultaneously and arterial access is therefore
needed. It is also useful to obtain blood gas samples from ascending and
descending aorta. Large ducts (> 10 mm) may require occlusion balloon catheters.
Such ducts are seldom suited for coil occlusion. Little data is currently
available on how balloon occlusion data can be interpreted in hypertensive
ducts. From our preliminary experience with 20 patients who had hypertensive
ducts, it appears that a decline in PA mean and PA diastolic pressures
to less than 25% of the baseline appears to predict a good long-term outcome
after duct closure.
Angiography
Aortography for profiling the PDA: In the event arterial access
is obtained a conventional aortogram in the left lateral view and 450
RAO views with the pig-tail catheter in the proximal descending aorta positioned
just distal to the ampulla usually allows reasonable definition of the
PDA. The use of a marker pigtail allows for accurate measurements of the
PA end of the duct. Large volumes (1-2 ml/Kg) of contrast medium at the
maximum possible flow rates recommended for the catheter (@
18 ml/sec) should be used.
Angiography when arterial access is not obtained
In our institution we do not routinely obtain arterial access for coil
occlusion of PDA in infants and small children.12 It is still
possible to obtain satisfactory an angiographic profile of the PDA without
arterial access using techniques outlined below.
After obtaining baseline hemodynamic information a long sheath passed via the femoral vein is positioned across the PDA in the descending aorta over a 0.038-inch guide-wire. The 45 cm Balkin contralateral introducer sheath (Cook Inc., Bloomington, IN) can be used for the PDA in infants and small children. This sheath has been originally designed for access to the contralateral femoral artery. This sheath has a shape that is well suited for the PDA and comes in sizes from 5.5F -7F. For small infants (< 3 Kg) with ducts < 3.5 mm a 4 F long sheath (> 25 cm long, Cook) can be used . When two or more 0.052-inch coils are to be simultaneously delivered, 7F contralateral or 8-9F Mullin’s sheath (Cook Inc.) is used. Once the Sheath is in the descending aorta, the dilator of the long sheath is removed. With the guide-wire in place, the sheath is withdrawn into the aortic end of the duct ampulla. A hand injection of 5 -10 ml of contrast into the sheath in the lateral and 450 right anterior oblique views usually allows excellent definition of the duct ampulla and measurement of duct diameter at pulmonary insertion. With guide wire across the sheath small adjustments in sheath position can be made for the best angiographic definition.
Measurements
The maximum diameter of the pulmonary artery (PA) end of the PDA should
be measured. The lateral view is usually better suited for this purpose
(Fig. 5).

The diameter of the pulmonary arterial end of the duct varies with the
cardiac cycle and the largest diameter should be identified through careful
frame-by-frame evaluation. Often the largest diameter is in a frame where
the duct is less well opacified. Ductal spasm is not infrequent in infants
and is provoked by attempts to cross it. (Fig. 6) In our institution we
prefer to use the echocardiographic measurement of the duct size to guide
the choice of coil diameter unless the angiographic measurement is larger.

Technique of bioptome assisted coil delivery
Coil selection
Gianturco coils with diameters at least twice the measured duct diameters
at PA end are chosen. We prefer to use the echocardiographic measurement
to guide coil selection unless the duct diameter by angiography is larger
than the echo measurement. This is done to avoid being misguided by duct
spasm. It is preferable to use simultaneously delivered multiple coils
for ducts that measure more than > 3 mm in diameter. This usually results
in a higher immediate occlusion rates. For ducts > 3.5 mm in diameter one
or more coils-should be 0.052-inch coils. For ducts > 4.5 mm in diameter
it may be necessary to deliver 3 coils simultaneously and for ducts > 6
mm in diameter four coils may have to be simultaneously delivered. Ducts
> 8 mm in diameter are best closed by the Amplatzer duct occluder. (Table
1).
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|

Deployment of coils (Fig. 8)
The coils are then delivered via the long sheath all the way and one
to two loops are extruded out of the tip of the sheath in the descending
thoracic aorta. The side arm of the sheath should be connected to the pressure
transducer. The entire assembly is then pulled back towards the pulmonary
artery until the sheath tip is just beyond the duct ampulla. Here coils
are almost entirely brought out of the sheath. Typically the retrograde
flow across the duct pushes the entire coil mass into the ampulla. There
is an abrupt cessation of oscillatory movements (as a result of pulsatile
forward flow in the descending aorta) together with compaction of the coils
as they enter the duct ampulla. All coil turns should compact in the ampulla.
This is sometimes difficult to ensure when coil diameters are very large
or when more than 4 coils are simultaneously delivered. The tracheal air
shadow can serve as an additional landmark to guide the coil placement.
The relationship of the tracheal air shadow to the duct ampulla and the
pulmonary arterial end of the duct is previously identified in the ductal
angiogram. Once the coils compact in the ampulla the sheath is slowly pulled
back until the pressure recorded from the side arm of the sheath declines
indicating that the tip of the sheath is now on the pulmonary artery. At
this point the bioptome is slowly pulled back until a small part of the
coil protrudes into the pulmonary artery. Some resistance is felt at this
stage. There is further compaction of the coil turns in the ampulla. Efforts
should be made, as far as possible, to leave less that half a turn to protrude
into the pulmonary artery. Contrast injection through the side arm can
be made to ensure that the coils are correctly positioned with free flow
into the branch pulmonary arteries. The jaw of the bioptome should be released
soon after a satisfactory position is obtained. Attempts to hold the coils
for longer periods should be avoided because this often results in one
or more coil-turns being inadvertently pulled into the pulmonary artery.
In the event that the coils chosen are too small for the duct, one or more
turns or the entire the coil mass can be pulled into the pulmonary artery
by the bioptome. The coils can be withdrawn into the sheath and redeployed
after addition of another larger coil to the coil mass. An echocardiogram
or aortogram (if arterial access was obtained) should be performed after
3 minutes. Additional 0.038-inch coils are delivered if a well-defined
jet of residual flow was demonstrable by either color Doppler or
angiography (see below). Small diffuse whiffs of flow often disappear over
the next 12-24 hours.

Delivery of additional coils (Fig.9)
Additional coils are contemplated whenever there is an unacceptable
amount of residual flow (clearly defined color jet on Doppler, filling
of the entire MPA on angiography). It is advisable to obtain arterial access
in the event the duct needs to be re-crossed. A 4 French introducer sheath
can be used for this purpose. Heparinization is needed at this stage. The
duct is crossed form the arterial end using a 0.035-inch glide wire (Terumo
or Roadrunner, Cook) and a 4 French right coronary catheter. Dacron fibers
of the previously deployed coils can get entangled if a regular Teflon
wire is used. Once the wire is across the duct the catheter should be gently
advanced over the wire. In general, 5 mm diameter, 5cm long coils work
well as additional coils in these situations.
Half a turn of the coil is brought out of the tip of the 4 F catheter
positioned in the MPA across the duct. The catheter is slowly pulled back
until the protruding half turn reaches the PA end of the duct. The coil
is then delivered by slowly withdrawing the catheter while keeping the
guidewire in the catheter in close contact with the coil. Catheter withdrawal
tends to pull the protruding coil out of the MPA and guidewire advancement
may result in excessive coil protrusion. Coil delivery has to be accomplished
in small steps ensuring at all times that the coil length protruding out
into the MPA is kept constant at half a turn until a substantial length
of coil is exposed out of the catheter (Fig. 9). Advancing the catheter
towards the ampulla usually results in formation of coil turns that should
be positioned in the ampulla. Hand injections of contrast 3 minutes after
coil delivery should be made to assess residual flows.

Complications
Embolization to the Branch Pulmonary Arteries (Fig. 10)
This can happen soon after coil release or rarely within 24 hours or
exceptionally after that. The dislodged coils usually embolize to the proximal
right or left pulmonary arteries if they are large and if multiple coils
are used. Single coils usually embolize distally to smaller branches. Sufficient
time is available for planned retrieval because instability is rare. The
long sheath should be retained in the MPA. A 4 French multipurpose catheter
or the 4F snare catheter should be passed via the long sheath and positioned
near the embolized coil mass with the help of a glide wire guided by the
movement of the coil when it comes in contact with the wire. An Amplatz
gooseneck snare (5 mm for children and small vessel embolization, 10 mm
for other situations) should be used to grasp the coil tip. When multiple
coils have been used for PDA closure it is important to hold the coils
at the sutured end. The coils must be captured into the long sheath in
the pulmonary artery because it is important to prevent the coil mass to
be entangled in the tricuspid valve tensor apparatus (Fig. 10).

Embolization to the descending thoracic aorta (Fig. 11)
When coil(s) embolize into the aorta, the duct should be immediately
re-crossed with a 5F multipurpose catheter or snare catheter. A 10 mm Amplatz
gooseneck snare (Microvena, MN, USA) should be used to hold the end of
the coil(s) and the same coil(s) can deployed in the duct once again as
the catheter is pulled back towards the MPA.

Loss of grip on the coil mass
The jaws of the bioptome may sometimes lose their grip on the coil
mass when coils are being pulled back into the long sheath after an initial
unsatisfactory deployment. A variable part of the coil remains in the sheath.
Attempts to recapture the coils with the bioptome or a snare are unlikely
to succeed and the coils may get pushed out of the sheath. A 3 F vascular
retrieval forceps (Cook) works well in this situation. The tip of the vascular
retrieval forceps has a short (3 cm), soft guidewire that can be positioned
adjacent to the coil tip in the sheath. The jaws of the forceps open adequately
enough to grasp the coil tip and retrieve the coil mass.
Inability to release the coil after bioptome jaws are opened
Occasionally coil tip remains in the jaws after they are opened. The
coils can be released by slow rotation the bioptome with the jaws open.
Alternatively, advancing the long sheath to the jaws of the bioptome helps
in the release of the coil.
Hemolysis from residual flows
Hemolysis is a rare but a serious complication of coil occlusion.17-18
For hemolysis to occur there often has to be clearly defined residual flow
at the end of the procedure together with an audible murmur. Of 1299 patients
who underwent PDA coil occlusion at our institution between August 1998
and June 2007, 7 patients (age, 6 weeks – 64 years) developed overt hemolysis.
All had large ducts and residual flows after the procedure. The occurrence
of hemolysis correlated significantly with both age as well as duct size.
Hemolysis was associated with a fall in hemoglobin of 3–6 g/100 ml (3 patients),
jaundice (3), and renal failure (1). Hemolysis subsided spontaneously in
two patients and additional coils had to be deployed in 3 patients. Once
hemolysis is established it is often difficult to eliminate flows and many
additional coils may be required. It is therefore important to be aggressive
and early intervention should be considered if residual flows are significant.
In our series, one 6-week-old infant continued to have significant flows
and ongoing hemolysis after three additional coils were deployed in two
settings. The hemolysis (and residual flows) in this infant only resolved
after exchange transfusion.
Post-procedure management
The patient can be sent home 6-8 hours after the procedure once recovery
from sedation or anesthesia is complete especially if arterial access has
not been used.15 For children in whom arterial access was obtained
we choose to keep them overnight. We consider it mandatory to obtain an
echocardiogram just prior to discharge for residual flows across the PDA,
LPA turbulence and aortic flows and ventricular function. A small fraction
of patients develop varying degrees of left ventricular dysfunction immediately
after duct closure. We suggest antibiotic prophylaxis for endocarditis
be maintained for 6 months after the procedure. We recommend follow-up
echocardiography, 3 months after the procedure and yearly thereafter for
3 years after the procedure unless there are residual issues.
Long term concerns
Compatibility with magnetic resonance (MR) imaging
The conventional stainless steel coils are not MR compatible and likely
to produce artifacts during imaging. Most manufacturers have started to
make coils using materials that are MR compatible (MR –eye coils, Cook
inc.). The 0.052-inch coils are still made of stainless steel. This is
an important limitation that needs to be overcome by manufacturers in the
future.
Stenosis of the left pulmonary artery origin
It is important particularly in infants to avoid excess coil loop protrusion
into the left pulmonary artery. Both after initial deployment and during
follow-up the LPA flows should be carefully evaluated by color Doppler.
Residual flows and re-canalization
Residual flows at 24 hours tend to persist. In addition, in a small
proportion of completely occluded ducts (0.3% in our experience) may re-canalize.
The indication for repeat coil occlusion is not clear. We recommend coil
occlusion if a murmur is audible.
Results of coil occlusion – a single institution experience
At our institution, coil occlusion is the preferred mode for treatment
of PDA primarily because it is the least expensive of all methods available.
From August 1998 to June 2007, 1299 PDA coil occlusions have been attempted
at our institution. The bioptome assisted technique of multiple coils was
used in 455 patients. The median duct size was 3 mm (range 1.7-10 mm).
Coil occlusion was feasible in a wide range of age groups 6 days –65 years.
A substantial proportion of these patients were infants; 53% weighed ?
10 Kg and 8.3 % (108 patients) weighed ? 5 Kg. Four preterm infants (900
gms – 1.5 Kgs) underwent successful coil occlusion. The procedure was unsuccessful
in 12 patients (0.9%). Immediate residual flows were seen in 11.6% of the
entire series. At 3 months follow up, 4.7% of the patients recorded residual
flows and 2% acquired stenosis of the left pulmonary artery origin as defined
by new color Doppler flow turbulence at the origin of the left pulmonary
artery. Coil embolization occurred in 98 patients (7.5%); 44 of which occurred
to the aorta and 54 to the branch pulmonary arteries. The incidence of
coil embolization declined to 4% in the last 200 cases. Six infants underwent
coil occlusion while on a ventilator for heart failure and / or pneumonia.
In the entire series there was one procedure related mortality. In this
infant the coil mass embolized to the right pulmonary artery and could
not be retrieved. Cardiac surgery was performed to retrieve the coil but
the infant succumbed to refractory pulmonary hypertension and right lung
injury.
Cost comparisons revealed a substantial advantage of coil occlusion over device occlusion. The average cost of coil occlusion in our institution was 29% that of the device. It is possible that re-sterilization and reuse of the bioptome would contribute to this advantage. However, like in many institutions world over, we also re-sterilize and reuse the delivery cable and the sheaths of the occlusive device. The cost of the bioptome and the delivery system for the occlusive device are quite similar.
Conclusions
Coil occlusion of ducts > 3 mm is technically feasible and this is
a substantially less expensive alternative to occlusive devices. It is
especially advantageous in selected small infants and preterm newborns
(with suitable anatomy) because it can be accomplished through a smaller
introducer sheath. It is necessary to pay attention to specific anatomic
and technical details to ensure success.
References
R Krishna Kumar
Amrita Institute of Medical Sciences and Research Centre
Elamakkara PO
Kochi 682026, Kerala, India
Phone: 91-484-280-4001599
Fax: 91-484-2802020
rkrishnakumar@aims.amrita.edu
|
|
|
|
|
![]() |